Industrial Hygiene Surface Sampling Method Summary



Name(s) of Compound Teriparatide, Teriparatide acetate, PTH, Human PTH 1-34
CAS # 52232-67-4
Method Report Number SBM121-R1
Analytical Method EIA
Swabbing Media CleanTips swabs TX761, 4 swabs per sample; in polypropylene test tube with dual position cap
Swabbing Solution 5% Acetic acid/DI Water
Quantification Limit 50 ng/sample
Average Recovery from Stainless Steel – 82.4%
Storage Conditions At least 7 days at ambient temperature
At least 14 days at 2 – 8 °C
Shipping Requirements Ambient temperature, overnight
Date Issued 12/07/2006

If you wish to submit a sample, please follow the sample submission procedures found at safebridge.com/laboratory/sample-submission. SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.