Industrial Hygiene Surface Sampling Method Summary

Name(s) of Compound Tamoxifen Citrate
CAS # 10540-29-1
Method Report Number SBM095-R0
Analytical Method HPLC-UV
Swabbing Media CleanTips swabs, Large Alpha swab, cat. TX714A, TWTexwipe
Swabbing Solution Methanol
Quantification Limit 1μg/sample
Average Recovery from Stainless Steel – 94.0%
Counter Top - 89.1%
Concrete Floor - 67.3%
Storage Conditions At least 7 days at ambient temperature
At least 21 days at 2 – 8 °C, or ~ -20 °C
Shipping Requirements Ambient temperature, overnight
Date Issued 08/18/2005

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.