Industrial Hygiene Surface Sampling Method Summary



Name(s) of Compound Spinetoram, XDE 175 (Mixture of Spinetoram factor J and factor L)
CAS # 935545
Method Report Number SBM221–R0
Related Document(s) SBM219-R1
Analytical Method HPLC-UV
Swabbing Media 3 Large Alpha swabs, CleanTip swab, cat. TX714A, TWTexwipe
Swabbing Solution Methanol: DI water (80:20, v/v)
Quantification Limit 2μg/sample
Average Recovery from Stainless Steel - 84.9%
Polycarbonate - 78.1%
Polyethylene - 94.0%
Marble - 85.8%
Vinyl Tile - 87.5%
Detection Limit of Analytical Method 0.1 µg/mL
Storage Conditions At ambient temperature for at least 8 days
At 2 – 8 °C or ~ -20 °C for at least 3 weeks
Shipping Requirements Ambient temperature
Date Issued 04/03/2013

If you wish to submit a sample, please follow the sample submission procedures found at safebridge.com/laboratory/sample-submission. SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at safebridge.com.