Industrial Hygiene Air Sampling Method Summary

Name(s) of Compound Niraparib Tosylate
CAS # 1038915-64-8
Method Report Number SBM260-R0
Analytical Method HPLC-FL
Sampling Media 25 mm Teflon filter, 1.0 micron pore size, Millipore, cat.# FALP02500, support pad, SKC, cat.# 225-28, in 3-pc polypropylene cassette, SKC, cat. 225-8585
Sampling Rate 2 L/min
Recommended Minimum Sampling Volume 100 L
Quantification Limit 5 ng/sample
Detection Limit of Analytical Method 1 ng/mL
Storage Conditions At ambient temperature, at 2 - 8 °C or at ~ -20 °C for at least three weeks
Shipping Requirements Ambient temperature
Date Issued 02/13/2015

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.