Industrial Hygiene Air Sampling Method Summary



Name(s) of Compound Lactase, beta-Galactosidase from Aspergillus Oryzae
CAS # 9013-11-2
Method Report Number SBM203-R0
Analytical Method ELISA
Sampling Media 25 mm Teflon filter, 1 micron pore size, cellulose support pad, in 2-piece polypropylene cassette
Sampling Rate 2 L/min
Recommended Minimum Sampling Volume 50 L
Quantification Limit 0.3 ng/sample
Storage Conditions At least 11 days at ambient temperature
At least 24 days at 2 - 8 °C, or ~ -20 °C
Shipping Requirements On ice pack
Date Issued 04/18/2012

If you wish to submit a sample, please follow the sample submission procedures found at safebridge.com/laboratory/sample-submission. SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at safebridge.com.

<< go back