Industrial Hygiene Air Sampling Method Summary

Name(s) of Compound Irinotecan HCl M1, SPT1025M1
CAS # 136572-09-3
Method Report Number SBM198-R1
Analytical Method HPLC-FL
Sampling Media 25 mm Glass fiber filter, type AE, support pad, in 3-pc polypropylene cassette
Sampling Rate 2 L/min
Recommended Minimum Sampling Volume 50 L
Quantification Limit 5 ng/sample
Storage Conditions At ambient temperature for at least 7 days
At 2 – 8 °C or ~ –20 °C for at least 21 days
Protected from heat, light, and moisture
Shipping Requirements Ambient temperature, protected from light
Date Issued 08/23/2012

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at