Industrial Hygiene Surface Sampling Method Summary

Name(s) of Compound Ganetespib, STA-9090
CAS # 888216-25-9
Method Report Number SBM230–R0
Analytical Method HPLC-UV
Swabbing Media Two CleanTips swabs, TX715
Swabbing Solution Methanol
Quantification Limit 50 ng/sample
Average Recovery from Glass - 97.8%
Stainless Steel - 97.8%
Lab Bench - 94.8%
Storage Conditions At ambient temperature for at least 8 days
At 2 – 8 °C or ~ -20 °C for at least 3 weeks
Protected from light
Shipping Requirements Ambient temperature, protected from light
Date Issued 10/22/2013

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.