Industrial Hygiene Air Sampling Method Summary



Name(s) of Compound DMx and conjugated DMx (cDMx)
CAS # N/A
Method Report Number SBM262-R1
Analytical Method ELISA
Sampling Media 25 mm Teflon, 1 micron pore size; support pad; in 3-pc polypropylene cassette
Sampling Rate 3 L/min
Recommended Minimum Sampling Volume 250 L
Quantification Limit 4 / 10 ng/sample
Storage Conditions At least 7 days at 2 - 8 °C
Shipping Requirements On ice pack, overnight
Date Issued 07/09/2015

If you wish to submit a sample, please follow the sample submission procedures found at safebridge.com/laboratory/sample-submission. SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at safebridge.com.