Industrial Hygiene Air Sampling Method Summary

Name(s) of Compound Cytarabine
CAS # 147-94-4
Method Report Number SBM155–R0
Analytical Method HPLC-UV
Sampling Media 25 mm PVDF filter, 5 micron pore size, no support pad, in 2-piece polystyrene cassette
Sampling Rate 2.5 L/min
Recommended Minimum Sampling Volume 50 L
Quantification Limit 10 ng/sample
Storage Conditions At ambient temperature for at least 7 days
At ~ -20 ºC for at least 14 days
Shipping Requirements Ambient temperature
Date Issued 11/17/2008

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at