Industrial Hygiene Air Sampling Method Summary

Name(s) of Compound Cyclosporin A; CSA
CAS # 59865-13-3
Method Report Number SBM259-R0
Related Document(s) SBM079-R0
Analytical Method HPLC-UV
Sampling Media 25 mmTeflon filter, 1 micron pore size, (Millipore cat No FALP02500) placed in cassette filter assembly (SKC cat No 225-71A) and IOM sampler (SKC No 225-70A)
Sampling Rate 2 L/min
Recommended Minimum Sampling Volume ≥ 45 L
Quantification Limit 90 ng/sample
Detection Limit of Analytical Method ~25 ng/mL
Storage Conditions At ambient temperature for at least 7 days
At 2 - 8 ºC or ~ -20 ºC for at least 21 days
Shipping Requirements Ambient temperature, protected from light
Date Issued 01/15/2015

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.

SafeBridge Consultants, Inc. is the premier occupational health and safety and risk assessment consulting firm to the pharmaceutical, biotechnology, and fine chemical industry. Learn more at