Industrial Hygiene Surface Sampling Method Summary

Name(s) of Compound Abarelix, Plenaxis, PPI-149, PPI-149-CMC
CAS # 183552-38-7
Method Report Number SBM039-A-R0
Analytical Method ELISA
Swabbing Media Two 37 mm glass fiber filters, 1 micron pore size in glass test tube with Teflon lined screw cap
Swabbing Solution 0.1% TFA/DI Water
Quantification Limit 2.5 ng/sample
Average Recovery from Stainless Steel - 93.1%
Aluminum - 86.0%
Plastic Caps - 87.4%
Rubber Stoppers - 89.5%
Glove Box Sleeve - 86.9%
Storage Conditions At least 42 days at 2 - 8 °C
Shipping Requirements Ambient temperature
Date Issued 05/01/2003

If you wish to submit a sample, please follow the sample submission procedures found at SafeBridge regularly develops new and custom industrial hygiene air and surface sampling methods. For more information, please call (650) 961-4820.